Preview

Different methods of correcting the readings of the DVGN-01 albedo neutron dosimeter placed behind the IBR-2M protection

https://doi.org/10.29317/ejpfm.2018020402

Abstract

The paper presents the results of correction of readings of DVBN-01 albedo dosimeters behind IBR-2M protection, the Neutron Physics Laboratory of the Joint Institute for Nuclear Research (JINR), Dubna, byvarious methods. The neutron spectra were measured at two points behind the IBR-2M protection in the experimental halls, and measurements were carried out with the spherical albedo system at these points. The correction coefficients for DVGN-01 were calculated from the measured spectra based on the data of the spherical albedo system. A good agreement of the coefficients calculated by different methods is shown, which indicates the reliability of the obtained values for the correction coefficients. Based on the results of this work and the data obtained in the other investigations, the values of correction factors
recommended for use in individual radiation control (IRC) in the FLNP were obtained.

About the Authors

Yu. V. Mokrov
Joint Institute for Nuclear Research
Russian Federation


S. V. Morozova
Joint Institute for Nuclear Research
Russian Federation


G. N. Timoshenko
Joint Institute for Nuclear Research
Russian Federation


V. A. Krylov
Joint Institute for Nuclear Research
Russian Federation


O. B. Badunov
Joint Institute for Nuclear Research
Russian Federation


G. A. Seilkhanova
Joint Institute for Nuclear Research
Russian Federation


References

1. Normy radiacionnoj bezopasnosti NRB-99/2009 (SanPiN 2.6.1.2523-09, Moskva, 2009). (In Russian)

2. Kompleks avtomatizirovannyj individual’nogo dozimetricheskogo kontrolja AKIDK-301 (Rukovodstvo po jekspluatacii, Angarsk, 2007). (In Russian)

3. Sannikov A.V. et al., Preprint IFVE-6 ORI (2005). (In Russian)

4. L.G. Beskrovnaya et al., Physics of Particles and Nuclei Letters 7(3) (2010) 212-221.

5. A.V. Sannikov et al., Preprint IFVE 2008-1 (2008) ORI. (In Russian)

6. Yu.V. Mokrov, S.V. Morozova, Physics of Particles and Nuclei Letters 11(2) (2014) 127-136.

7. Yu.V. Mokrov et al., Physics of Particles and Nuclei Letters 11(6) (2014) 809-817.


Review

For citations:


Mokrov Yu.V., Morozova S.V., Timoshenko G.N., Krylov V.A., Badunov O.B., Seilkhanova G.A. Different methods of correcting the readings of the DVGN-01 albedo neutron dosimeter placed behind the IBR-2M protection. Eurasian Journal of Physics and Functional Materials. 2018;2(4):307-314. https://doi.org/10.29317/ejpfm.2018020402

Views: 402


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 2522-9869 (Print)
ISSN 2616-8537 (Online)