•  
  •  
 

Abstract

The paper presents the results of correction of readings of DVBN-01 albedo dosimeters behind IBR-2M protection, the Neutron Physics Laboratory of the Joint Institute for Nuclear Research (JINR), Dubna, byvarious methods. The neutron spectra were measured at two points behind the IBR-2M protection in the experimental halls, and measurements were carried out with the spherical albedo system at these points. The correction coefficients for DVGN-01 were calculated from the measured spectra based on the data of the spherical albedo system. A good agreement of the coefficients calculated by different methods is shown, which indicates the reliability of the obtained values for the correction coefficients. Based on the results of this work and the data obtained in the other investigations, the values of correction factors
recommended for use in individual radiation control (IRC) in the FLNP were obtained.

First Page

307

Last Page

314

Creative Commons License

Creative Commons Attribution 4.0 International License
This work is licensed under a Creative Commons Attribution 4.0 International License.

Share

COinS